Settings

Parameter

Description

Mode

Range

CPAP

AutoSet

AutoSet for Her

Mode

Sets the therapy mode available on the device.

Set Pressure

Sets the fixed treatment pressure.

4-20 cm H2O (4-20 hPa), 0.2 cm H2O (0.2 hPa) increments

Pressure Range

Sets the pressure range for treatment

Select Min Pres: 4-20  cm H2O (4-20 hPa), 0.2 cm H2O (0.2 hPa) increments. Select Max Pres: 4-20 cm H2O (4-20 hPa) 0.2 cm H2O (0.2 hPa) increments. Dependend on Min Pres.

Parameter

Description

Mode

Range

S

ST

T

VAuto

Mode

Sets the therapy mode available on the device.

IPAP

Sets the pressure to be delivered to the patient when the device is triggered into inspiration.

4-25 cm H2O (4-25 hPa), 0.2 cm H2O (0.2 hPa) increments

EPAP

Sets the pressure to be delivered to the patient when the device is cycled into expiration.

S, ST, T: 3–25 cm H2O (3–25 hPa) ASV: 4 - 15 cm H2O 0.2 cm H2O (0.2 hPa) increments

Max IPAP

Sets the maximum IPAP (inspiratory pressure) delivered by the device.

4-25 cm H2O (4-25 hPa), 0.2 cm H2O (0.2 hPa) increments.

Min EPAP

Sets the minimum EPAP (expiratory pressure) delivered by the device.

4-25 cm H2O (4-25 hPa), 0.2 cm H2O (0.2 hPa) increments.

Pressure Support (PS)

Difference between IPAP and EPAP. Adjust for patient comfort.

0-10 cm H2O (0-10 hPa), 0.2 cm H2O (0.2 hPa) increments.

Respiratory Rate

Sets the breaths per minute (BPM) or 'backup' rate.

5 - 50 BPM

Backup Rate

Sets the breaths per minute (BPM) or 'backup' rate.

5 - 50 BPM

Easy-Breathe

Enable/Disable the Easy-Breathe feature.

On / Off

Ti Range

Set the minimum/maximum limit on the time the device spends in IPAP.

Min: 0.1-4.0 sec, 0.1 sec increments. Max: 0.3-4.0 sec, 0.1 sec increments.

Ti

Sets the duration of inspiration in timed breath. Dependent on Respiratory Rate.

0.3–4.0 sec, 0.1 sec increments

I:E

Ratio of inspiratory duration to the expiratory duration.

Rise Time

Sets the time taken for pressure to increase from EPAP to IPAP. The Rise Time scale can be approximately read as 'milliseconds' (e.g., 200 is approximately 200ms).

Min/ 150-900ms, 50ms increments.

Trigger

Sets the level of inspiratory flow above which the device changes from EPAP to IPAP.

Very Low/ Low/ Med/ High/ Very High

Cycle

Sets the level of inspiratory flow below which the device changes from IPAP to EPAP.

Very Low/ Low/ Med/ High/ Very High

Parameter

Description

Mode

Range

ASV

ASVAuto

Mode

Sets the therapy mode available on the device.

EPAP

Sets the pressure to be delivered to the patient when the device is cycled into expiration.

S, ST, T: 3–25 cm H2O (3–25 hPa) ASV: 4 - 15 cm H2O 0.2 cm H2O (0.2 hPa) increments

EPAP Range

Sets the minimum and maximum EPAP delivered by the device.

Min / Max EPAP range 4-15 cm H2O (4-15 hPa) Max EPAP is dependent on Min EPAP level.

PS (Pressure Support) Range

Sets the pressure support range to be delivered by the device.

Min PS Range:

0-6 cm H2O (0-6 hPa)

Dependent on Maximum operating pressure and EPAP level (in ASVAuto mode, dependent on Min EPAP)

Max PS Range:

5-20 cm H2O (5-20 hPa)

Dependent on Min PS Range. Upper limit dependent on Maximum operating pressure and EPAP level.

Parameter

Description

Mode

Range

CPAP

AutoSet

AutoSet for Her

AutoSet Response

Set to Standard or Soft. If Soft is selected, patients will receive gentler pressure rises during therapy.

Standard/Soft

Ramp Time

Ramp setting (minutes)

Off/ 5-45 mins

Start Pressure

Set the pressure at the start of ramp, up to treatment pressure.

4-Set pressure, 0.2 cm H2O (0.2 hPa) increments

EPR

Enable/Disable EPR

On/ Off

EPR Type

Available when EPR is enabled.

Full Time/ Ramp Only

EPR Level

Set the EPR value

1/2/3 cm H20 (1/2/3 hPa)

Climate Control

Climate Control setting.

Manual/Auto

Tube Temperature

Heated tube setting (°C / °F)

Off/ 60–86°F (16–30°C), 2°F (1°C) increments / Auto (only available when Climate Control is set to Auto)

Humidity Level

Humidity setting.

Off/ 1-8

Parameter

Description

Mode

Range

S

ST

T

VAuto

Ramp Time

Ramp setting (minutes)

Off/ 5-45 mins

Start EPAP

Set the pressure at start of ramp, up to minimum treatment pressure.

S, ST, T:

3*-EPAP, 0.2 cm H2O (0. 2 hPa) increments.

VAuto/ASV:

4-EPAP, 0.2 cm H2O (0. 2 hPa) increments.

ASVAuto:

4-Min EPAP, 0.2 cm H2O (0. 2 hPa) increments.

*Dependent on minimum operating pressure.

Climate Control

Climate Control setting.

Manual/Auto

Tube Temperature

Heated tube setting (°C / °F)

Off/ 60–86°F (16–30°C), 2°F (1°C) increments / Auto (only available when Climate Control is set to Auto)

Humidity Level

Humidity setting.

Off/ 1-8

Parameter

Description

Mode

Range

ASV

ASVAuto

Ramp Time

Ramp setting (minutes)

Off/ 5-45 mins

Start EPAP

Set the pressure at start of ramp, up to minimum treatment pressure.

S, ST, T:

3*-EPAP, 0.2 cm H2O (0. 2 hPa) increments.

VAuto/ASV:

4-EPAP, 0.2 cm H2O (0. 2 hPa) increments.

ASVAuto:

4-Min EPAP, 0.2 cm H2O (0. 2 hPa) increments.

*Dependent on minimum operating pressure.

Climate Control

Climate Control setting.

Manual/Auto

Tube Temperature

Heated tube setting (°C / °F)

Off/ 60–86°F (16–30°C), 2°F (1°C) increments / Auto (only available when Climate Control is set to Auto)

Humidity Level

Humidity setting.

Off/ 1-8

Parameter

Description

Range

Patient View

Set the level of access available to the patient.

Simple/ Advanced

SmartStart™

Enable/disable the SmartStart feature. If you enable the SmartStart feature, the device will start automatically when the patient breathes into the mask.

On/ Off

SmartStop

Enable/disable the SmartStop feature. If you enable the SmartStop feature, the device will stop automatically when the patient removes the mask.

On/ Off

Reminders

Mask

Set a recurring reminder to replace the mask.

Off/ 1/ 3/ 6/ 9/ Yearly

Tube

Set a recurring reminder to replace the air tubing

Off/ 1/ 3/ 6/ 9/ Yearly

Filter

Set a recurring reminder to replace the air filter

Off/ 1/ 3/ 6/ 9/ Yearly

Humidifier

Set a recurring reminder to replace the humidifier tub

Off/ 1/ 3/ 6/ 9/ Yearly

Parameter

Description

Selection

Language

Set the display language

(Not all languages available in all regions)

English / Français / Español / Português

TimeZone

Set up the correct time zone

GMT time zone

Temperature Units

Set the temperature units

°F / °C

Restore Defaults

Reset to default settings (except for language, date and time)

OK to restore defaults. Cancel to return to previous menu.

Erase data

Erase all data stored on the device and SD card. Settings, date, time and device run hours are not affected.

OK to erase data

Cancel to return to previous menu.

About

View Run Hours, SW version, CF version, network provider, network type and signal strength (dBm) of the device.

Parameter

Description

Range

Mask Setting

Select the type of mask used by the patient. Refer to Mask Device Compatibility List on www.resmed.com.

Full Face / Nasal / Pillows

Tube

Select the type of air tubing used by the patient. ClimateLineAir air tubing is automatically detected when connected to the device.

SlimLine / Standard / 3m

BV filter

Select On if you attach a bacterial/viral filter.

On/Off