Skip to main content

Sleep Lab Help

Technical specifications

90W AC adaptor

Dimensions (H x W x L)

1.3" x 2.6" x 5.2" (32 mm x 65 mm x 132 mm)

Weight

0.55 lb (250 g)

AC input range

100–240 V, 50–60 Hz 115 V, 400 Hz

DC output

24 V

Class of equipment

Class II

TxLink 2

Dimensions (H x W x L)

1.8" x 4.9" x 5.8" (45 mm x 125 mm x 147 mm)

Weight

0.77 ± 0.11 lb (350 ± 50 g)

TxLink 2

90W AC adaptor (with ResMed therapy device attached)

Typical power consumption

3 W

77 W

Peak power consumption

5 W

101.5 W

Housing construction

Flame retardant engineering thermoplastic

Operating temperature

+41o F to +95o F (+5oC to +35oC)

Operating humidity

10%–95% non-condensing

Storage and transport temperature

-13o F to 158o F (-25oC to 70o C)

Storage and transport humidity

5% to 95% non-condensing

Operating altitude

Sea level to 9,870 ft (3,010 m)

Analog outputs

6 channels

All channels isolated from mains

3 kV

All channels isolated from communications signals

1.5 kV

No isolation between channels

-

Voltage range

-1 VDC to +1 VDC

Accuracy

± 5%

Bandwidth

25 Hz

Output impedance

50 Ohm

Electromagnetic compatibility:

Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2 (class B). Information regarding the electromagnetic emissions and immunity of these ResMed devices can be found on ResMed.com/downloads/devices.

IEC60601-1 classification

Class II

Design life

5 years

Intended operator

Sleep technician or clinician

Notes

  • The manufacturer reserves the right to change these specifications without notice.

  • The performance of the analog outputs is determined by the attached ResMed therapy device.