Troubleshooting

Problem/Possible cause

Solution

TxLink 2 fails to power up

Incorrect power supply

Check the ResMed supplied power supply is connected correctly.

Bad connection from power supply to TxLink 2 DC input

Faulty mains power source

Check the mains power source and associated cabling.

PAP device fails to power up

Incorrect TxLink 2 PAP device cable

Check TxLink 2 Power LED is On.

Bad connection from TxLink 2 DC output to PAP device 2 DC input

Check the ResMed supplied TxLink 2 PAP device cable is connected correctly.

TxLink 2 Error LED is ON

Incorrect power supply

Check the ResMed supplied power supply is connected correctly.

TxLink 2 device damaged

Check the TxLink 2 for signs of physical damage.

No devices found displayed on Administrator Add tab

TxLink 2 not powered up

Check the TxLink 2 power LED is On and network LED is flashing.

TxLink 2 not connected to Ethernet network

Check Ethernet cabling if network LED is Off.

TxLink 2 IP address configuration settings are incorrect

Check the TxLink 2 IP address configuration settings with your network administrator.

Use the form on the Add tab to manually add the TxLink 2 IP address.

TxLink 2 name displays in red on Connection manager

TxLink 2 has lost power

Check the TxLink 2 power LED is On and network LED is flashing.

TxLink 2 has lost network connection

Check Ethernet cabling if network LED is Off.

Cannot connect to PAP device

PAP device is not powered up

Check TxLink 2 PAP device LED is On.

Check the ResMed supplied TxLink 2 PAP device cable is connected correctly.

Access code requested but not known

TxLink 2 has a prior secure connection with another EasyCare Tx 2 instance

Refer to System Administration (see System Administration) for how to display the access code.

TxLink 2 in use

Another EasyCare Tx 2 instance is already connected to the TxLink 2.

Close the titration window on the other EasyCare Tx 2 instance.

Access code did not work when making a secure connection to a TxLink 2

Access code timed out

Access codes will expire after 5 minutes. Add the access code within 5 minutes.

Titration window is open on the PC that generated the access code

Close the titration window on the PC that generated the access code. Generate a new code and repeat the connection process.

Note

For any serious incidents that occur in relation to this device, these should be reported to ResMed and the competent authority in your country.