Rise time adjustment
Available in: S, ST and T mode only.
Rise Time sets the time taken for the device to reach IPAP. The greater the rise time value, the longer it takes for pressure to increase from EPAP to IPAP.
Patients with a high ventilatory demand may prefer a shorter rise time, while patients who are slow breathers may prefer a longer rise time.
Note
A prolonged rise time inhibits fast pressurization, therefore, rise time should not be set longer than Ti Min or the patient's normal inspiratory time.