Skip to main content

Sleep Lab Help

Min/Max Pressure Support

ASV and ASVAuto Pressure Support (PS) defines the difference between peak inspiratory pressure and minimum expiratory pressure. ASVand ASVAuto modes automatically adjust pressure support between Max PS and Min PS to maintain the patient's ventilation target. The pressure support level must have sufficient travel between these settings to allow the algorithm to respond to variations in patient sleep patterns.