Skip to main content

Sleep Lab Help

Troubleshooting

If there is a problem, try the following suggestions. If you are not able to fix the problem, contact your local ResMed dealer or ResMed office. Do not open the device.

General troubleshooting

Problem/possible cause

Solution

Air is leaking from around the mask

Mask may be fitted incorrectly.

Make sure the mask is fitted correctly. See the mask user guide for fitting instructions or run the Mask Fit function.

The patient is getting a dry or blocked nose

Humidity level may be set too low.

Increase the Humidity Level.

There are droplets of water in the mask and air tubing

Humidity level may be set too high.

Decrease the Humidity Level.

Tube temperature may be too low

Increase the Tube temp

The patient is getting a very dry mouth

Air may be escaping through the patient's mouth.

Increase the Humidity Level.

The patient may need a chin strap to keep the mouth closed or a full face mask.

The patient feels that too much air is being delivered from the device

Ramp may be turned off

Use the Ramp Time option.

The patient feels that not enough air is being delivered from the device

Ramp may be in progress

Wait for air pressure to build up or turn Ramp Time off

Ramp start pressure may be too low

Increase Ramp start pressure.

No display

Backlight on the screen may have turned off. It turns off automatically after a short period of time

Press the Start therapy /standby button located at the top of the device or touch the screen.

Power may not be connected.

Connect the AC adaptor and make sure the plug is fully inserted.

Therapy has stopped but the device is still blowing air

Device is cooling down

Device blows a small amount of air in order to avoid condensation in the air tubing. It will stop automatically after 30 minutes.

Humidifier tub is leaking

Humidifier tub may not be assembled correctly.

Check for damage and reassemble the humidifier tub correctly.

Humidifier tub may be damaged or cracked.

Replace the humidifier tub.

The patient is not getting enough air/oxygen flow is disrupted

Tubing or humidifier tub may be blocked

Check for blockages. Reconnect the tubing and reassemble the humidifier tub correctly.

SmartStart is enabled, but the device does not automatically start when the patient breathes into their mask

Breath is not deep enough to trigger SmartStart

To start therapy, take a deep breath in and out through the mask, before breathing normally.

Press the Start therapy/Standby button located on the top of the device

There is excessive leak

Adjust the mask and headgear

Air Tubing may not be connected properly. Connect firmly at both ends.

SmartStop is enabled but does not automatically stop when the patient removes their mask

Incompatible mask being used

Only use equipment recommended by ResMed.

Contact ResMed or see ResMed.com for more information

If the patient is using a conduit mask, SmartStop will not work.

Device error messages

Device message/possible cause

Solution

High leak detected. Check your humidifier or side cover

Humidifier tub may not be inserted properly.

Make sure the humidifier tub is correctly inserted.

High leak detected. Connect your tubing

Air tubing may not be connected properly.

Make sure the air tubing is firmly connected at both ends.

Mask may be fitted incorrectly.

Make sure the mask is fitted correctly. See the mask user guide for fitting instructions or use the Mask Fit function to check the mask fit and seal.

Tubing blocked. Check your tubing

Air tubing may be blocked.

Check the air tubing and remove any blockages. Press the Start therapy/Standby button to restart the device.

Read only card. Remove, unlock and re-insert SD card

SD card switch may be in the lock (read-only) position.

SD card functionality is not intended to be used for the AirCurve 11 VPAP Tx.

System fault, refer to user guide, Error 4

Device may have been left in a hot environment.

Allow to cool before re-use. Disconnect the AC adaptor and then reconnect it to restart the device.

Air filter may be blocked.

Check the air filter and replace it if there are any blockages. Disconnect the AC adaptor and then reconnect it to restart the device.

There may be water in the air tubing.

Empty the water from the air tubing. Disconnect the AC adaptor and then reconnect it to restart the device.

All other error messages, for example, System Fault, refer to user guide, Error X

An error has occurred on the device.

Remove power and restart device. If error persists, contact your local ResMed dealer or ResMed office. Do not open the device.